Medical Machines/Microtomes/Remove cowling
Appearance
Removing the cowling
[edit | edit source]-
1. Support the left side of the machine to reach four M5 screws. Two fasten the side cover and two fasten the left side of the middle cover.
-
2. Remove the screw fastening the side cover at the rear of the machine.
-
3. Remove the screw fastening the middle cover at the left rear corner.
-
4. Remove the screw fastening the side cover at the front of the machine.
-
5. Remove the screw fastening the middle cover at the left front corner.
-
6. Remove support block and left cover.
-
7. Support the right side of the machine to reach four M5 screws. Two fasten the side cover and two fasten the right side of the middle cover.
-
8. Remove the M5 screw fastening the side cover at the rear of the machine.
-
9. Remove the M5 screw fastening the middle cover at the right rear corner.
-
10. Remove the M5 screw fastening the side cover at the front of the machine.
-
11. Remove the M5 screw fastening the middle cover at the right front corner.
-
12. Remove the support block. Remove the section thickness knob and shaft by pulling out. The shaft is held in the dial by a spring loaded ball.
-
13. Remove right cover.
-
14. Lift off the middle cover.